How to do rotary evaporator troubleshooting?
Rotary evaporator
is widely used to remove solvent in cbd oil distillation process, and economical and idea instrument for concentration, purification, evaporation, etc. -
How to do simple troubleshooting by self when you meet the common problems ofrotary evaporator?
1. The motor of rotary evaporator does not turn
– Connect the broken line and replug.
– Update the fuse or confirm that there is no abnormality in the power supply.
– Update the circuit board or electrical control box.
– Connect the ground wire or change the work location.
– Exclude according to the inverter manual P33.
2. Rotavapor’s water bath / oil bath is not heated
– Detect the power supply is correct.
– Change the heating coil. Replace the solid state relay or relay board.
– Change the temperature controller. Detect the probe wiring or update the probe.
3. Bath pot temperature detection
– Change the temperature controller.
– Replace the solid state relay or relay board.
– Change the probe or temperature controller of vacuum rotary evaporator
.4. Vacuum can’t pump
– Empty the solvent and try the empty flask of rotary evaporators
.– Vacuum oil pump for oil (water) changing, cleaning and maintenance.
– Detect and eliminate piece by piece along the vacuum line.
– Reassemble, clean the glass wool sandwich panel, apply vacuum grease, and tighten the flanges.
-
If your problems of rotary evaporator have not included in above situations, or you need some spare parts of rota evaporator, please feel free contact [email protected]
to get technical support from rotary evaporator manufactures
directly.
Comments (2)
I have a problem, My rotary evaporator is flushing water back from the pump (water-based pump) into the condenser and is collected in the solvent collector flash (back flushing), how do i solve this problem, please i need help here.
Hi Ben,
We will get back to your Email shortly.
Thanks