How To Install Laboratory Spray Dryer? TOPTION Spray Dryer Supplier
TOPTION brand laboratory spray dryer is widely used to the translation of liquid material to trace power, there’s no need to filtrate, concentrate, and disrupt before dry, and has wide application for all aqueous solutions such like emulsion and suspension. -
The main features of lab spray dryer:
Air Inlet Temperature Range
0 to 300℃
Air Outlet Temperature Range
0 to 120℃
Evaporation
3000 ml/H
Electric Heating Power
3KW/AC220V/Single Phase Ground Line
Fan
0.2KW
Max Airflow
5.5 m³/min
Max Air Pressure
686Pa
COMPA Working Pressure
2-5Bar
Overall Dimensions
780*620*1600(mm)L*W*H
Diameter of Spray Nozzle
0.75mm
-
How to install spray drying machine?
1. Hold up the drying chamber and fit it to the fixed pallet (place in the white PTPE cushion block, and then lock the quincunx handle).2. The cyclone separator fits directly into the exhaust stack after the nut, sealing ring & stainless steel shim fit into the air-outlet tube of cyclone separator, adjust the positions of air-outlet of the drying chamber & air inlet of cyclone separator, until they can be put in order completely, and lightly tighten the nut.3. Join sample collection flask and cyclone separator with Clip hoop.4. Join the waste collection tube and the main chamber with Clip hoop.5. Sprayer fits into the equipment and fit the 4mm nylon tube(for de-block piston) and 6mm nylon tube(for spray).6. The silicone tube is fitted through the peristaltic pump and placed into the sample feed inlet of sprayer. -